Target Product Profile: enzyme Immunoassay (EIA) Antibody tests to help determine if people have antibodies to SARS-CoV-2
Updated 8 August 2023
1. Target Product Profile
1.1 Enzyme Immunoassay (EIA) Antibody tests to help determine if people have antibodies to SARS-CoV-2
Issued by MHRA
1.2 Version Control
- 1.0 Initial document
1.3 The purpose of a Target Product Profile “TPP”
Target product profiles (TPP) outline the desired ‘profile’ or characteristics of a target product that is aimed at a particular disease or diseases. TPPs state intended use, target populations and other desired attributes of products, including safety and performance-related characteristics. They help guide industry development towards desired characteristics. A TPP provides a common foundation for the development of tests that contains sufficient detail to allow device developers and key stakeholders to understand the characteristics a test must have to be successful for the particular intended use. Included is a description of
(1) the preferred and
(2) the minimally acceptable profiles based on the intended use, setting of use, and intended user, with respect to the performance and operational characteristics expected of the target products.
1.4 TPPs for COVID-19
These product profiles have been developed to assist manufacturers to design and deliver tests that might be useful in support of Pillar 3 of the UK testing strategy. The TPP assists the UK government in making decisions regarding central procurement of antibody tests and might also be used in local procurement decisions. Any deviation from existing standards must be fully justified. Production lead time will also factor into decision making.
Implementation of Pillar 3 of the testing strategy relies on availability of antibody tests that could tell people whether they have had the virus. Such tests usually require taking a blood sample and looking for the presence of antibodies specific to SARS-CoV-2, the causative agent of COVID-19.
A positive result from this test does not guarantee immunity to COVID-19 infection and may not indicate the infectious status of the person. A negative test does not guarantee no prior COVID-19 infection. Some people may not develop an antibody response.
1.5 Clinical performance requirements
This is a specification of the clinically acceptable specifications for a laboratory Enzyme Immunoassay (EIA) test to be made and used in the UK during the current COVID-19 pandemic caused by SARS-CoV-2 virus. It sets out the clinical requirements based on the consensus of what is ‘minimally acceptable’ in the opinion of UK IVD industry, healthcare professionals and medical device regulators given the emergency situation. A test kit with other specifications than this may not be suitable to support Pillar 3 of the UK testing strategy.
The intended use of assays that match these profiles (or one that does not yet meet the specifications but looks promising) is to determine if an individual has previously been exposed to SARS-CoV-2 (and not other coronaviruses circulating in the population).
The criteria for clinical specificity is set deliberately high in a test intended to detect an antibody response where the result may be given in the absence of a known past PCR (Polymerase chain reaction) positive result. In a test with low specificity, there is an unacceptable risk that a person is incorrectly told that they have made an antibody response. They may consequently be exposed to infection and be at risk of illness and may also pass that infection on to others that they come in contact with. This is particularly a concern in people from a high-risk group, or a group which isdirectly exposed to vulnerable persons.
Manufacturers claiming to have an assay intended to show clinical immunity, or that correlates with potential immune protection, must have scientific evidence to support the claim. They should also have made a risk assessment in line with ISO14971 for medical devices to address the intentional and unintentional use of the results of an antibody test.
It should be noted that a different TPP may be required for different case uses. Other future use cases may need to consider, for example, use of home collection of samples such as capillary blood or saliva with appropriate validation of performance and usability; or samples from hospitalised convalescent patients.
As such the contents of the TPPs in this document are restricted to those supporting use for Pillar 3. Failure to meet the criteria set out in the TPP does not necessarily mean that a test doesn’t have wider applications for use in the UK*. These TPPs are profiles based on our best information, but the science is rapidly evolving. The TPP is subject to review and may need to be updated at short notice.
1.6 Key
Acceptable: defines the minimum acceptable feature
Desired: highly desirable features of considerable benefit. As time is of the essence if omitting one of these features significantly accelerates development and production it should be considered.
2. TPP COVID-19 serology (antibody) Enzyme Immunoassay (EIA) test
2.1 Scope
Intended use
To determine if an individual has made an antibody response to SARS-CoV-2.
Evidence for each intended use must include that which supports the specificity of the immune response to SARS-CoV-2 (acceptable feature) and if claim is of immunity (desired feature), that the antibodies detected are protective.
Each intended use must have evidence for scientific validity and include that which supports the specificity of the immune response to SARSCoV-2 Assays intended to show an immune response that correlates with potential immune protection must either explicitly measure neutralising antibodies or measure binding antibodies that have a proven correlation with neutralising titre.
Target population
Desired: to determine possible immune protection against SARS-CoV-2
Acceptable: people who may have recovered from suspected or confirmed SARS CoV-2 infection or may have previously developed an asymptomatic infection
Target user
Laboratory trained Health care professionals
Target use setting
Medical laboratories working to appropriate quality & competence standards, e.g. ISO15189:2012
2.2 Test design characteristics
Test format
Either:
- a kit containing all materials required for testing. This includes a standard, coated 96 well EIA plate, and all necessary reagents including, (as needed) specimen diluent, wash buffer, conjugate, substrate, acid, calibrators (if required) and controls and their requisite dilution material
OR
- a kit containing all lot specific material required for testing. This includes a standard, coated 96 well EIA plate, and all related lot-specific reagents such as conjugate and calibrators (if required). Supply of any non-lot specific material (eg controls, wash buffer, diluents, substrate and acid etc) provided separately
Specimen collection requirements, unless specific to the assay, do not need to be addressed.
Note: assay may be automated. Multi-test formats other than 96 well EIA plates, using different technologies, may be useful.
Target analyte
Desired: IgG antibodies to SARSCoV-2 virus
Acceptable: total antibodies (IgG or IgA or IgM) to the SARS-CoV-2
Note: design should incorporate a target which is correlated with previous infection or immune protection The kinetics of the humoral response for COVID-19 are not yet fully understood but total antibody (IgA, IgM, and IgG in any combination) may be useful.
Sample type
Whole blood &/or plasma, &/or serum.
If more than one sample type is specified clinical sensitivity and specificity must be determined for each claimed sample type.
Result output
Desired: qualitative/semi-qualitative
Acceptable: qualitative
Where quantitation (full or semi) is the output, the interpretation of the result must be clear and linked, with evidence. Lot to lot precision must be considered in assigning any such quantitation.
Note: international/national reference standards used when available.
Assay controls
Controls that provide evidence that the IVD has functioned correctly – they must accurately monitor performance of the assay against the critical performance claims.
Assays compatible with external quality assessment schemes desirable.
Identification capability
Ability to link patient/donor identification must be feasible.
For plate format this must include identification system of wells or sample positions or possibility to barcode.
Pack size
Single or multiple test kits e.g. 96 wells EIA plates
Need for calibration/spare parts
Calibrators and User serviceable spare parts should be available if required for the assay system
2.3 Performance characteristics
Clinical sensitivity
Greater than 98% (with 95% confidence intervals of 96-100%) on specimens collected 20 days or more after the appearance of first symptoms.
These statistics rely on testing of at least 200 confirmed positive cases.
Note: see Introduction - clinical performance requirements.
Clinical specificity
Greater than 98% (within 95% confidence intervals 96-100%).
These statistics rely on testing of at least 200 confirmed negative cases or from testing of specimens collected at least 6 months before the known appearance of the virus.
Note: see Introduction clinical performance requirements.
Analytical specificity
Desired: no known cross-reactivity with other known coronavirus, common respiratory pathogens
Acceptable: minimal cross-reactivity with other known coronavirus, or common respiratory pathogens
Refer to list in annex for relevant pathogens. Known Cross reactions should be listed in the IFU.
2.4 Test procedure characteristics
Sample preparation
Some processing acceptable with standard laboratory equipment.
Need to process sample prior to performing test. Appropriate containment environment must be specified.
Specimen volume in assay
Desired: 5-50uL
Acceptable: 5-50uL
Can be prediluted.
Reagent volume per well
Desired: volume of reagents to well (apart from wash) less than 100uL.
Acceptable: volume of reagents to well (apart from wash) less than 100uL.
Volumes apply to standard 96 well EIA plate formats.
Result
Comes with either easy manual calculations or a software programme for result calculations and monitoring of QC.
Specify requirements for the assay, e.g. plate reader, automated plate washer, spectrophotometer.
Biosafety
No additional biosafety should be needed to use of standard medical laboratory practice
2.5 Operational characteristics
Test kit storage conditions
Desired: 5 – 30 ⁰C, 80% relative humidity
Acceptable: 2 – 8⁰C
Operating conditions
Either 15 - 25⁰C or 37⁰C for specimen and conjugate incubations
Kit reagent stability
At least 6 months at 2-8⁰C or 12 months at -200⁰C.
Accelerated stability testing is acceptable provided it is supported by real time stability studies.
In use stability
All reagents (as presented) stable for up to 12 test runs over a period of a minimum of 3 months
End point stability (time window during which signal remains valid)
If not automated, greater than 30 minutes.
Fully automated system may read signal as part of the process.
Disposable requirements
Device and accessories should be disposed in standard biological waste containers, no glassware Or be biodegradable or combustible.
Training needs (Time dedicated to training session for end users)
For both: no special training requirements for laboratories familiar with performing routine manual or automated EIAs
2.6 Other
Result interpretation
Clear instructions for grey zones, interpretation of quantitative or semiquantitative results, including significance when comparing results taken in time sequence (i.e. comparing results taken on different days). Requirements for confirmation or repeat testing clearly described. All result interpretation must be supported by evidence.
Instructions for use (IFU)
In line with IVDD (98/79/EC) Annex 1 requirements:
-
simple interpretation to aid sampling and results interpretation and what to do with the test if the control fails
-
clear reading time
-
instructions for results interpretation across reporting range
-
clear warnings of limitations for use including expected performance characteristic
*paper or electronic
Interferences
Interferents should be included in risk evaluation from endogenous and exogenous sources.
See Annex for possible examples.
Hook effect
Desired: assay design is such that potential for false results due to a hook effect is not an issue
Acceptable: assays should specify their linear reporting range and upper limit of reporting
Definition: The phenomenon whereby the effectiveness of antibodies/antigens to form immune complexes is sometimes impaired when concentrations of an antibody or an antigen are very high.
Regulatory status
CE marked, or in process of meeting EU regulatory requirements for in vitro diagnostic medical devices
Design and manufacturing environment
Conforms to:
-
BS EN ISO 14971:2019 Medical devices. Application of risk management to medical devices
-
ISO 13485:2016 Medical devices. Quality management systems, Requirements for regulatory purposes
Labelling and IFU
In accordance with Annex I of the IVD Directive under essential requirements.
Compliant with IVD Medical Device Directive.
3. Annex: Assay validation
3.1 Establishing performance characteristics
The following aspects should be considered when designing and validating the assay:
-
When available, reference material should be used to establish performance, including seroconversion panels, quality control materials and proficiency testing materials
-
There is no currently agreed reference standard for establishing specimen immunity status. In the absence of such agreed position, it is recommended that a reference standard used for establishing truth is a composite standard, comprised of the following: “Appropriately timed specimens collected from symptomatic patients diagnosed in a laboratory with validated assays.” Technical documentation should include your rationale for your specimen characterisation and also any discrepant result analysis.
-
When establishing analytical specificity, the following should be considered:
- prepandemic samples (specimens collected at least 6 months before the known appearance of the SARSCoV-2 virus)
- other coronavirus, SARS-CoV-2
- hCOV 229E, OC43, HKU1, NL63 epitopes
- Adenovirus (e.g. C1 Ad. 71)
- Human metapneumovirus (hMPV)
- Parainfluenza virus 1 – 4
- Influenza A & B
- Enterovirus (e.g. EV68)
- Respiratory syncytial virus
- Rhinovirus
- Chlamydia pneumoniae
- Haemonphilus infuenzae
- Legionella pneumophila
- Mycobacterium tuberculosis
- Streptococcus pneumoniae
- Streptococcus pyogenes
- Bordetella pertussis
- Mycoplasma pneumoniae
- Pneumocystis jirovecii (PJP)
Potential interferents may originate from the following endogenous and exogenous sources:
Endogenous substances
- Haemoglobin
- Bilirubin
- Protein
- Triglycerides
- Rheumatoid Factor
- Total IgM
- Polyclonal hypergammaglobulinemia
- Hematocrit
- Antibodies developed against protein expression system used to generate recombinant antigens where relevant
- Heterophiles
Exogenous substances
- Recommended anticoagulants e.g. EDTA
3.2 Other
Confirmation tests: consider reference laboratory PCR or reference serology algorithm.