Target Product Profile: Laboratory-Based SARS-CoV-2 Viral Detection tests
Updated 8 August 2023
Version control
Version | Date issued | Description |
---|---|---|
1.0 | October 2020 | Initial document |
2.0 | 10 November 2021 | Changes to Clinical Sensitivity and Limits of Detection to accommodate assays with additional benefits, including: multiplex, short turn around times or high throughput. |
Increased Clinical Specificity in line with the changing state of the art and continuous quality improvement | ||
Updated regulatory sections in line with the UK departure from the EU | ||
Updated intended-use and included statements concerning SARS-CoV-2 variants of concern | ||
Updated sections on Reference Standard and Sample types | ||
Harmonised introductory sections and Annex’s with newer TPP formatting | ||
Included considerations of clinical performance evaluation, clinical impact, downstream sequencing and carry over | ||
Removed requirements considering supply volumes, equipment size, sample volume |
The purpose of a Target Product Profile (TPP)
MHRA Target Product Profiles (TPP) are guidance documents which aim to support and accelerate the development and evaluation of new medical technologies to address specific unmet clinical or public health needs of high strategic priority to the UK population.
TPPs summarise the key features and anticipated performance specifications of a new device in advance, to enable innovators to design and develop high quality products that are fit for purpose and meet specific health-related goals. They are intended to be used to support product design, research and development planning and to facilitate discussions with regulators.
MHRA TPP’s are aspirational documents aimed at test manufacturers and are based upon the best available evidence and independent expert opinion. They do not represent UK government policy and are not regulatory requirements. For information on the current National technical validation process and their relevant performance goals, please see National technical validation process for manufacturers of SARS-CoV-2 (COVID-19) tests - GOV.UK (www.gov.uk)
MHRA TPPs are living documents that are reviewed on a frequent basis, dependent on the specific disease area, and updated as additional evidence and information becomes available. Manufacturers should ensure they are working to the most recent version of a TPP.
Intended use for this TPP
This TPP is intended to be used by manufacturers to support the development of laboratory-based in vitro diagnostic tests for detection of current SARS-CoV-2 infection in people with and without symptoms., as part of UK government national testing programmes or accredited private testing services.
The exact performance requirements will vary depending on the specific use-case, taking into consideration the following:
- The target population (e.g. Children, young people, adults, older people)
- The setting (e.g. A&E, testing centres, workplaces)
- The intervention decision being informed (e.g. release from isolation, contact tracing, infection control measures etc).
This TPP does not consider tests for use in the following possible scenarios:
- Self-Tests for use at home
- Point of care tests to be performed by a professional (see here for full definitions) in a health and/or social care setting.
- Prognose a patient’s likely outcome, including disease severity or survival.
- Predict or monitor a patient’s likely response to treatment.
Clinical performance requirements
The TPP sets out the requirements based on the consensus of what is “desired” and minimally “acceptable” in the opinion of healthcare professionals and scientists given the current situation. Products meeting the “desirable” criteria will likely have a role in a greater number of use-cases than products that only meet the “acceptable” criteria.
The decision to use a particular test for a specific use must be informed by clinical expert opinion at the time, considering the disease prevalence, risks, benefits and downstream consequences of testing vs not-testing. Annex 2 provides tables and further discussion which may be useful in supporting decision making, by demonstrating the impact of changing sensitivity, specificity and prevalence on the numbers of false positives and negatives.
Future developments
These profiles are subject to review and change, as we gain a greater knowledge of the virus, the disease and our needs for an effective response. They may need to be updated at short notice.
As our knowledge and understanding of the disease changes and the UK clinical needs change, so will the specifications. A test that meets this version of the TPP may not meet future versions.
Key to Table
Desired: Highly desirable features of considerable benefit. As time is of the essence, if omitting one of these features significantly accelerates development and production it can be considered.
Acceptable: Defines the minimum acceptable feature.
Target Product Profile: COVID-19 - Laboratory-based SARS-CoV-2 Viral detection testing
Scope
Key feature
Intended Function(s)
Desired
Used in the detection of current SARS-CoV-2 infection
And
Differentiation from recent or recovered infection.
And/or
Differentiation of SARS-CoV-2 from other respiratory infections.
And/or
Differentiation of SARS-CoV-2 Variants of Concern (VoCs)
Acceptable
Used in the detection of current SARS-CoV-2 infection
Comments
Current infection: an infection in which the causative organism has the potential, either now or in the future, to cause disease or onward transmission. An individual with a current infection may not display disease symptoms, require treatment or be infectious at the time of testing.
Manufacturers should indicate if the test is only intended for use in a specific population or setting and would be expected to justify the balance between risk and benefits. Tests with “acceptable” levels of sensitivity and specificity may only have application in a limited number of use-cases and will likely require additional risk mitigation measures, such as confirmatory testing or infection control. Refer to Annex 2.
Whilst multiplexing is desirable for specific use-cases, it may not be appropriate for all.
Target population
Adults and children with or without symptoms
Comments
Populations “without symptoms” represents a range of pre-symptomatic, peri-symptomatic, sub-symptomatic and truly asymptomatic phenotypes, spanning all stages of disease (early and late) and a wide range of viral loads (low to high).
Manufacturers should carefully consider the design features of tests intended for use in children to ensure their safe and effective operation and acceptability.
Manufacturers should ensure that risk-management and clinical performance is appropriate for their claimed target populations. Evidence of clinical performance should be provided for each population claimed by the manufacturer and performed by the relevant user group.
As the immune status of the UK population changes over time, manufacturers should undertake studies to evaluate how the test continues to perform in vaccinated and re-infected individuals.
Target user
Trained healthcare / public health professionals (i.e. one of the 10 health and social care professional bodies that are overseen by the professional standards authority) and suitably trained and assessed as competent lab technicians or scientists.
Comments
A target user will perform the assay, interpret and communicate the results.
Full training appropriate to the intended user is required.
Target use Setting
Healthcare and Medical Laboratories
Comments
These exclude Point of Care (POC) testing environments which will utilise tests types that are appropriate for that setting.
Includes both private and public sector laboratories.
Test design characteristics
Test format
A standardised kit that contains all materials required for the laboratory procedure that includes controls, reagents and Instruction for Use (IFU).
All equipment and accessories needed to perform the assay and sample processing included, with the exception of routine laboratory consumables such as pipettes, pipette tips, disinfectants, screw cap eppendorfs, heat blocks, fluorescent readers etc.
Should the technology not be compatible with routinely used sample collection devices these should also be provided at the same time as the test kit.
Comments
May apply to, for example:
- Open and Closed high throughput platforms
- Microarray testing (for broad range pathogen testing)
All accessories need to be validated for use in combination with the test as part of the CE, CE UKNI or UKCA marking.
Specifications for general reagents that can be used with open or closed systems must be clearly defined.
Target Analyte (Measurand)
Desired
Dual (or more) SARS-CoV-2 targets (e.g. nucleic acid(s), antigen(s) or other targets)
Single (or more) targets for a range of infectious respiratory viruses, including influenza A virus, and influenza B virus, RSV A & B
Single (or more) targets for all current SARS-CoV-2 variants of concern
Acceptable
Single SARS-CoV-2 target (e.g. nucleic acid, antigen or other target)
Comments
Evidence demonstrating the association of the analyte(s) with current SARS-CoV-2 infection (scientific validity) should be provided.
Manufacturers should consider targeting assays to conserved regions/epitopes of SARS-CoV-2 to ensure the detection of current and future variants
Dual antigen targets could include multiple epitopes for the same protein.
Multiplex sytems must be able to clearly distinguish between targets included in the panel.
Assays that detect whole SARS-CoV-2 Virus may also be acceptable
There is a requirement on suppliers to confirm performance in detection of both current and emerging strain variants of SARS-CoV-2 as they arise and confirm this to MHRA where VOC (variants of concern) or VUI (variants under investigation) are reported in line with MHRA requirements, see Annex 1
Sample type
Desired
Method not requiring a swab (e.g. saliva, sputum, stool, breath sample).
Validation of assays for use with respiratory tract samples’ (sputum, endotracheal, bronchoalveolar lavage, nasopharyngeal aspirate), tissue samples or cerebral spinal fluid may also be desirable for some specific use cases.
Acceptable
Nasal and/or throat swab
Comments
All sample types claimed as appropriate in the instructions for use must be validated as part of the performance assessment. Study sample sizes for each sample type must be sufficiently powered.
Methods not using invasive swabs are desirable due to the individual’s discomfort and pre-analytical errors. They may also facilitate improved quantification of viral load.
Manufacturers should ensure that swabs are appropriate for the intended population (i.e. smaller swabs for children).Consideration should be given to compatibility of sample with collection media/buffers and assay performance
Not all sample collection buffers or methods are compatible with downstream testing requirements e.g. genotyping and sequencing.
Compatible sample collection
Desired
Test is validated for use with the sample types listed above which may be collected into a variety of sample collection medias e.g. viral transport medium, inactivation medium, solvents, dry swabs, and saline in addtional to media provided with the kit.
Sample media is suitable for downstream processing including sequencing for identification of lineage, mutations and variants of concern
Acceptable
Test is validated for use with sample types above, which may be collected into a media provided with the kit.
Comments
Additional processing or re-sampling may be required if the availability of quality RNA for downstream processing cannot be met by any single test method
Result output
Desired
Semi-Quantitative
Acceptable
Qualitative
Comments
Whilst it is desirable for analytical methods to be quantitative, the nature of swab sample collection means that results are likely to only be semi-quantitative.
For semi-quantitative assays, an appropriate number of calibration points and replicates covering the range of reliable signal should be applied.
Qualitative results may not support all downstream applications (e.g sequencing).
Power requirement
Desired
Standard mains power supply with UPS and the capability for battery power.
Acceptable
Mains power supply
Internal controls
Desired
Whole process positive controls, negative controls, internal and external controls are required to confirm validity of end-end processing and clearly identify sample inadequacy results as invalid.
Acceptable
Should include positive controls and negative controls with option to include internal controls/external controls in a single or multiplex format.
Comments
Invalid results may be due to sampling technique, the presence of biological inhibitors or matrix effects.
Technical failure rate
Desired
Less than 0.2%
Acceptable
Less than 1%
Comments
In use failures resulting from mechanical, controls, calibration or other factors which may not be regarded as reportable as an adverse event under the UK Medical Devices Regulations 2002 (SI 2002 No 618, as amended) (UK MDR 2002).
Does not include failure due to sample collection or technical issues outside of scope of test (e.g house-keeping genes)
Ease of use and result interpretation
Suitable for target user groups (i.e trained healthcare professionals)
Need for calibration
Desired
No calibration required
Acceptable
Remote or auto-calibration
Identification capability
Desired
Unique barcode or equivalent for integration into electronic systems
Acceptable
Labelling of the device with the subjects identification must be feasible
Carry Over
Bioinformatics approaches used to detect human-human sample contamination to monitor carryover.
Comments
The risk of carryover should be evaluated at each step of the assay.
Performance characteristics
Clinical (diagnostic) sensitivity (or Positive Percent Agreement)
Desired
Greater than 99% (with 95% two-sided confidence interval entirely above 97%).
Acceptable
Greater than 80% (with 95% two-sided confidence interval entirely above 70%).
Comments
At least 150 positive cases (per sample type). Greater sample numbers will be required to support higher performance claims.
Multiplex assays and assays with shorter turn-around-times or greater throughput may justify the lower 80% sensitivity requirement, in specific use cases. However, standard RNA extraction based RT-PCR assays should be achieving the Desirable performance.
Clinical (diagnostic) specificity (or Negative Percent Agreement)
Desired
Greater than 99% (with 95% two-sided confidence interval entirely above 97%).
Acceptable
Greater than 97% (with 95% two-sided confidence interval entirely above 93%).
Comments
A minimum of 250 COVID-19 negative cases (per sample type).
Testing should include all claimed specimen types and provide details of collection device and transport medium that have been validated for use with the assay.
Clinical Performance Evaluation
Desired
Positive and negative cases should be recruited prospectively and consecutively or randomly from the target population without prior knowledge of their disease status (e.g. single-gate design).
Studies directly comparing a new assay to the current state of the art test would be advantageous.
Test samples should be collected in the target setting at the same time as samples for the reference standard and any comparator methods.
In some cases the type of test or order of testing may need to be randomised.
Alternative study designs looking at the impact of using tests on disease spread (for tests to rule-in) or outbreaks (for tests to rule-out) should also be considered.
Acceptable
When prevalence is low, or in emergency use situations (e.g. a surge in cases of a new variant of Concern (VoC)), case-control designs using clinical samples with viral load distributions (determined by PCR) generalisable to the target population may be necessary.
Once in clinical use, manufacturers should immediately start to collect and make available in a timely way “Desirable” clinical performance evidence from post-market surveillance studies.
Comments
Two-gated case-control designs can introduce selection and spectrum bias and should be avoided if possible, refer to Rutjes, 2005. Claims made using such approaches (minimum criteria) are likely to overestimate the Clinical Performance.
Testing of all claimed specimen types should be performed with sufficient power, and details provided for collection devices and transport media that have been validated for use with the assay.
Practical recommendations for designing diagnostic accuracy studies in low prevalence settings can be found in Holtman, 2020.
Further information on the design and conduct of clinical performance evaluations of SARs-CoV-2 tests can be found in Doust, 2021.
Reporting of clinical performance evaluation studies should be in line with STARD 2015. Results should include a diagram of participant flow, participant clinical and demographic characteristics including distibutions of disease severity/stage and alternative diagnoses, time intervals between index and reference tests,and a 2x2 table of results in addition to the measures of diagnostic performance with 95% confidence intervals.
All efforts should be made to establish the disease/infection status of a study participant. E.g. evidence of prior infection (describing test results and timing) and vaccination status (date of vaccination/s, vaccine).
Clinical Reference Standard
Desired
A composite clinical reference standard, against which the clinical sensitivity and specificity are calculated.
This could, if scientifically valid and appropriate for the defined context of use, include considerations of immunity status, disease phase, virus characteristics, and infectivity.
Acceptable
A validated CE, CE UKNI or UKCA marked RNA extracted RT-PCR laboratory method in current clinical use that itself performs within the desirable analytical and clinical performance specifications of this TPP, against which the Negative/Positive Percent Agreement is calculated.
Comments
See the NICE evidence standards framework for more information on composite clinical reference standards.
An example of a temporary Composite Reference Standard for COVID-19 can be found here.
For samples with discordant results further testing could be done to try and explain the direction of discordance (for example, repeating the sample run on both tests or using a third method, if available). But this should not influence claims of sensitivity and specificity.
Analytical specificity
Desired
No clinically relevant cross reactivity or interference to all organisms and agents listed in Annex 1.
Inclusive of all SARS-CoV-2 variants of concern. Demonstrated in silico and in vitro, where suitable reference materials are available.
Acceptable
No clinically relevant cross reactivity to common seasonal respiratory pathogens.
Minimal interference caused by common interferents at clinically relevant concentrations (dependant on sample type and analyte)
Inclusive of SARS-CoV-2 variants of concern. Demonstrated in silico and in vitro, where suitable reference materials are available.
Comments
Manufacturers should consider inclusivity, exclusivity, cross-reactions and exogenous/endogenous interference. See annex 1 for list.
In silico analyses including database searches should be performed to confirm the species specificity, avoiding the possibility of accidental cross reaction with human and bacteria.
Limit of Detection (LOD)
Desired
An appropriate unit of measurement for the target analyte (e.g. International Units) equivalent to a viral load of less than 100 SARS-CoV-2 RNA copies/mL of sample.
Standard RNA extraction based RT-PCR assays should be achieving the Desirable LoD performance.
Acceptable
An appropriate unit of measurement for the target analyte (e.g. International Units) equivalent to a viral load of less than 10,000 SARS-CoV-2 RNA copies/mL of sample.
Multiplex assays and assays with shorter turn-around-times or greater throughput may justify the acceptable LoD performance.
Comments
The LoD is the lowest concentration of analyte that can be consistently detected in ≥95% of samples tested under routine laboratory conditions and in the appropriate sample matrix. This concentration must yield an assay value that can be reproducibly distinguished from values obtained with samples that do not contain the analyte. Refer to appropriate standards (e.g. CLSI EP17) in the design of studies. Ideally multiple batches/lots of kits/reagents should be used when establishing the LoD.
Where an appropriate International Standard, reference material or reference measurement procedure is available for the analyte(s) this should be used. For examples see here.
If there are no comparative reference materials or measurement procedures available, evidence should be provided to demonstrate the choice of strategy for determining LOD is appropriate.
To demonstrate equivalence of the analyte(s) with viral load in copies/mL, the quantity value and measurement uncertainty of the clinical samples used should be assigned using an appropriate reference method (e.g. dPCR). Commercially available quality control materials may not be value assigned with sufficeint accuracy to enable LoD evaluation.
The evaluation of LoD for some analytes using samples characterised in terms of RNA copies/mL may not be optimal and different criteria may need to be considered, if accompanied by sufficient evidence of scientific validity.
Clinical impact
Desired
Evidence that the test improves system and/or an individual’s outcomes (for example, time to diagnosis, subjects experience, use of pre-cautionary COVID-19 isolation facilities). Evidence that the test provides good value.
Comments
Refer to NICE evidence standards for further information.
The purchase price/cost of a test alone may not be a good indicator of its valueNICE use the incremental cost effectiveness ratio (ICER) as their preferred measure of value. Whilst NICE do not have a fixed threshold for cost-effectiveness, technologies exceeding £30,000/QALY have a higher probability of rejection and need to identify a strong case with regards to the certainty of evidence and innovative nature of the technology.
Turnaround time
Less than 90 minutes from sample to result.
Acceptable
Less than five hours from sample to result.
Comments
The time is from receipt of sample in test laboratory.
Throughput
Desired
More than 200 tests per machine/module every 4 hours.
Acceptable
More than 300 tests per machine/module every 24 hours.
Test procedure characteristics
Hands-on time
Desired
Less than five minutes per sample.
Acceptable
Less than 20 minutes per sample.
Sample processing and handling
Standardised sample-processing steps, using medical laboratory equipment (centrifuge, vortex, pipette etc).
Biosafety
Desired
Buffers or other components provided with kit or sample collection devices are proven to inactivate SARS-CoV-2 and any other respiratory viruses in scope of assay enabling entire process to be carried at CL2.
Acceptable
Requires inactivation pre-step at CL3 or CL2+. Remaining steps can be preformed at CL2.
Comments
Systems specifying need for lower biosafety environments must demonstrate sample inactivation and virus containment. Instructions for use must confirm biosafety requirements for sample handling and inactivation evidence
Appropriate consideration should be given for biosafety in the environment which the test should be carried out.
Risk in use
Risks have been managed according to ISO 14971.
Operational characteristics
Test kit storage and stability conditions
Desired
No cold chain (15 to 30º C).
Acceptable
Storage of kit and reagents at -20ºC and above for at least 12 months. Stable for 3 hours once removed from cold storage.
Comments
Packaging to be as compact as possible to facilate storage. Must be made clear if reagents can withstand freeze and thaw and will not be detrimental to the assay.
Assay end point stability (time window during which signal remains valid)
Desired
Up to 1 hour.
Acceptable
Up to 30 minutes.
Operating conditions
15 to 30ºC.
Connectivity
Desired
Wireless and cable connectivity via LIMS systems.
Acceptable
Cable connectivity via LIMS system.
Comments
Results may need to be transferred by digital media e.g. CSV files.
Presentation of results
Desired
Easy to capture for interpretation and able to record public health data.
Results do not require post run analysis (automatically called). Ability to access raw data e.g. RT-qPCR/RT-LAMP fluorescence traces.
Acceptable
Easy to capture for interpretation and able to record public health data.
Comments
Integrity of data must be maintained.
Reproducibility
More than 95% between repeats at LoD. More than 99% at higher concentrations.
Comments
Manufacturers should consider ISO 20395:2019 and ISO 5725-1 when evaluating reproducibility.
Disposal requirements
No additional disposal requirements beyond normal laboratory practice. Any special containment or disposal requirements need to be clearly specified (e.g. contamination control for RT-LAMP post amplification products).
Training needs (Time dedicated to training session for end users)
Desired
Less then half day training needed for for laboratory staff.
Acceptable
Less than one day of training needed.
Other
Label and Instructions for Use
Conforms to UK MDR 2002 and relevant designated standards.
Regulatory status
Desired
CE, CE UKNI or UKCA marked.
Acceptable
Exempt according to Regulation 12 of the UK MDR 2002.
Comments
For further information on the regulation of medical devices on the UK market, please see our guidance.
Specific guidance for manufacturers of COVID-19 tests can be found here.
Maintenance
Desired
Preventive maintenance should not be needed until after 2 years or 100,000 samples. An alert should be included to indicate when maintenance is needed.
Acceptable
Preventive maintenance should not be needed until after 1 year or 10,000 samples; an alert should be included to indicate when maintenance is needed.
Comments
Assuming the equipment is used at capacity 24 hours a day, seven days a week.
Design and manufacturing environment
ISO 13485:2016
Annex 1: Assay validation
Establishing Performance Characteristics
It is recommended that the following aspects are considered when designing and validating the assay. Analytical performance evaluations should test any multiplex molecular test system in its final configuration, and not in separate singleplex experiments.
- Reference material should be used to establish performance, including standard validation panels, quality control materials and proficiency testing materials
- When establishing the performance of the test, manufacturers must consider the analytical sensitivity and specificity of the test to SARS-CoV-2 Variants Under Investigation (VUI) and of Concern (VOC) , including those listed on gov.uk by Public Health England). A full up to date list of variants can be found at https://www.gisaid.org/. Manufacturers are expected to routinely perform in silico analysis of listed VOC and VUI of gov.uk as a standard post market surveillance analysis. Where an assay is suspected or known to be affected by a listed VOC OR VUI, the manufacturer should inform MHRA within 48 hours of discovery.
- When establishing analytical specificity, the following should be considered:
- Samples from patients who have received any licenced vaccine at several time points post vaccination ( e.g. <5 days and 1-6 months).
- pre-pandemic samples,
- other coronavirus, SARS-CoV-1,
- MERS- coronavirus
- hCoV 229E, OC43, HKU1, NL63
- Adenovirus (e.g. C1 Ad. 71)
- Human Metapneumovirus (hMPV)
- Parainfluenza virus 1-4
- Influenza A & B
- Enterovirus (e.g. EV68)
- Respiratory syncytial virus
- Rhinovirus
- Chlamydia pneumoniae
- Haemophilus influenzae
- Legionella pneumophila
- Mycobacterium tuberculosis
- Streptococcus pneumoniae
- Streptococcus pyogenes
- Bordetella pertussis
- Mycoplasma pneumoniae
- Pneumocystis jirovecii (PJP)
- Potential interferents may originate from the following endogenous and exogenous sources and may be more relevant to ligand-binding based antigen tests than conventional PCR based assays. Manufacturers should declare if any other endogenous/ exogenous substances will impact the assay.
- Antibacterial, systemic
- Antibiotic, nasal ointment
- Anti-viral drugs
- Antibodies developed against protein expression system used to generate recombinant antigens
- Bilirubin
- Biotin
- Blood (human)
- Haemoglobin
- Human Anti-mouse Antibody (HAMA)
- Medications most often prescribed in the population for which the test is ordered
- Mucin: bovine submaxillary gland, type I-S
- Nasal sprays or drops
- Nasal corticosteroids
- Nasal gel
- Protein
- Rheumatoid Factor
- Throat lozenges, oral anaesthetic and analgesic
- Triglycerides
In addition the assay must deliver across the clinical range of haematocrit values.
Post Market Performance considerations
An increase in the number of cases with new SARS-CoV-2 variants has been observed in the United Kingdom. MHRA is aware that some laboratories have reported poor performance of some diagnostic assays that include an S-gene target. Such impact on test performance may be linked to the virus variant alpha or other variants. Mutations are not exclusive to the S-gene and action is required by manufacturers regardless of the diagnostic assay targets.
In line with UK MDR 2002 vigilance and field safety corrective action reporting requirements, MHRA consider reports relating to Variants of Concern (VOC) to be serious public health threats, therefore significant safety issues should be reported within 48h.
Actions specific to new VOCs:
- Inform MHRA of the outcome of your initial risk assessment on the performance of your assay in light of identified variants of concern and your plan to mitigate against any new risks from mutations, including your timelines for addressing these.
- If the performance of your assay is directly impacted by new virus variant(s), a Field Safety Notice should be issued immediately to alert customers.
- A Post Market Surveillance plan (PMSP) should be in place to continuously monitor, investigate and assess newly emerging variants of SARS-CoV-2. The PMSP can include:
a. Fortnightly in silico checks of assay targets against GISAID sequence databases (https://www.gisaid.org) [*Please note that high profile potential issues should be immediately investigated] b. Scientific literature and post market intelligence gathering c. Outcomes of EQA schemes when available d. Use of reference materials when available e. Reporting potential safety issues of any new clinically significant variant SARS-CoV-2 strain on the performance of your assay to the MHRA
Public Visibility of device assurance
The MHRA intend to publish safety actions resulting from manufacturer in silico analysis and in vitro testing against variants of concern. The manufacturer should request if specific submitted information should not be made public. Public access to test device assurance will mutually benefit commercial suppliers and test device users.
Annex 2: Diagnostic accuracy considerations
When considering procurement and deployment of devices for any given clinical use-case, it is recommended to consider the maximum number of false positives and false negatives (Table 1) that would be acceptable for the new test based on the possible consequences of these misdiagnoses. It is also helpful to consider the post-test probability that someone with a positive or negative result has infection (Table 2). Programmes and testing services deploying self-tests into routine use should consider such information when determining where and when testing may be clinically/cost-effective and acceptable to end users.
Table 1: The tables below presents the numbers of false positives and negatives in a cohort of fixed size (1,000,000) with varying prevalence of SARS-CoV-2 infection.
Prevalence 20%
Numbers per 10,000 tested
Sensitivity | |||||
---|---|---|---|---|---|
Specificity | Test result | 99% | 97% | 95% | 80% |
99% | False +ves | 8000 | 8000 | 8000 | 8000 |
False -ves | 2000 | 6000 | 10000 | 40000 | |
95% | False +ves | 40000 | 40000 | 40000 | 40000 |
False -ves | 2000 | 6000 | 10000 | 40000 |
Prevalence 5%
Numbers per 10,000 tested
Sensitivity | |||||
---|---|---|---|---|---|
Specificity | Test result | 99% | 97% | 95% | 80% |
99% | False +ves | 9500 | 9500 | 9500 | 9500 |
False -ves | 500 | 1500 | 2500 | 10000 | |
95% | False +ves | 47500 | 47500 | 47500 | 47500 |
False -ves | 500 | 1500 | 2500 | 10000 |
Prevalence 1%
Numbers per 10,000 tested
Sensitivity | |||||
---|---|---|---|---|---|
Specificity | Test result | 99% | 97% | 95% | 80% |
99% | False +ves | 9900 | 9900 | 9900 | 9900 |
False -ves | 100 | 300 | 500 | 2000 | |
95% | False +ves | 49500 | 49500 | 49500 | 49500 |
False -ves | 100 | 300 | 500 | 2000 |
Prevalence 0.2%
Numbers per 10,000 tested
Sensitivity | |||||
---|---|---|---|---|---|
Specificity | Test result | 99% | 97% | 95% | 80% |
99% | False +ves | 9980 | 9980 | 9980 | 9980 |
False -ves | 20 | 60 | 100 | 400 | |
95% | False +ves | 49900 | 49900 | 49900 | 49900 |
False -ves | 20 | 60 | 100 | 400 |
Table 2: The following tables show the post-test probability of having an infection as the prevalence (or pre-test probability) changes. The percentage of people testing positive who are infected (Positive Predictive Value)) and the percentage of people testing negative who are infected (1-Negative Predictive Value) are shown for different prevalences from 0.2% to 20% (1 in 500 to 1 in 5). Programmes and testing services should determine their own acceptance criteria.
Desirable Test
Sensitivity 9% Specificity 99%
Prevalence | 0.2% | 0.5% | 1% | 2% | 5% | 10% | 20% |
---|---|---|---|---|---|---|---|
% +ves infected (PPV) | 16.56% | 33.22% | 50% | 66.89% | 83.9% | 91.67% | 96.12% |
% -ves infected (1-NPV) | 0.0020% | 0.0051% | 0.0101% | 0.0202% | 0.0505% | 0.1009% | 0.2016% |
Acceptable Test
Sensitivity 80% Specificity 95%
Prevalence | 0.2% | 0.5% | 1% | 2% | 5% | 10% | 20% |
---|---|---|---|---|---|---|---|
% +ves infected (PPV) | 3.11% | 7.44% | 13.91% | 24.62% | 45.71% | 64% | 80% |
% -ves infected (1-NPV) | 0.042% | 0.105% | 0.210% | 0.4193% | 1.0417% | 2.0619% | 4.0404% |
Annex 3 Glossary
- BSL: Biological Safety Level
- CL: Containment level
- dPCR: Digital polymerase chain reaction
- IVD: in-vitro diagnostic
- LAMP: Loop-mediated isothermal amplification
- LIMS: Laboratory Information Management System
- LOD: Limit of Detection
- NAT: Nucleic acid testing
- PCR: Polymerase chain reaction
- PPE: Personal Protective Equipment
- qPCR: Quantitative polymerase chain reaction
- RNA: Ribonucleic acid
- RT: Reverse transcription
- VOC: Varients of concern
Analytical sensitivity: quotient of the change in an indication and the corresponding change in the value of a quantity being measured (ISO 15193)
Analytical specificity: ability of a measurement procedure to determine solely the quantity it purports to measure (ISO 15193)
Clinical (Diagnostic) Sensitivity: ability of an IVD examination procedure to identify the presence of a target marker(s) in individuals with a particular disease or condition
Clinical (Diagnostic) Specificity: ability of an IVD examination procedure to recognise the absence of a target marker(s) in individuals with a particular disease or condition
Positive Percent Agreement: the proportion of non-reference standard positive subjects in whom the new test is positive.
Negative Percent Agreement: the proportion of non-reference standard negative subjects in whom the new test is negative
Positive Predictive Value: the proportion of patients with positive test results who have the target condition (as determined by the reference standard)
Negative Predictive Value: the proportion of patients with negative test results who do not have the target condition (as determined by the reference standard)